EKO LIDAR system is a Multi LIDAR System for the real time profiling
of "Temperature", "Humidity" and "Aerosol".
- EKO LIDAR system is an eye safe LIDAR using UV range laser of
the wavelength 355nm.
- EKO LIDAR system has been co-developed by EKO Instruments Co.,
Ltd. and Prof.Takao Kobayashi from the University of Fukui.
- EKO LIDAR system is useful for the meteorological and environmental
observation, and is used for seeking the solution for heat island
phenomenon and green house effect
|
Features
- EKO LIDAR system is designed with taking safety into consideration.
- EKO LIDAR system is placed in the air conditioned room. Therefore, the
laser equipment and optical parts can be used with high performance.
Measurement principle
| - Temperature |
Temperature profile is measured by detecting the Doppler dispersion
of Rayleigh scattering, which is changed according to the temperature.
|
| - Humidity |
Humidity profile is measured by detecting the Raman scattering intensity
of water vapor molecules. |
| - Aerosol |
Aerosol profile is measured by detecting the Mie scattering intensity. |
|
|
|
|
Mie scattering, Rayleigh scattering
|
Raman scattering
|
Measurement results
|
|
|
|
Temperature profile
(Height : 0 - 5000m,
color scale 0 - 30 deg.C)
|
Humidity profile
(Height : 0-2500m,
color scale 0 - 100%)
|
Measurement results
| Items |
Specifications |
| Laser |
Nd : YAG THG Third harmonic |
| Center wavelength 355nm (Eye safe) |
| 20Hz |
| Telescope |
Diameter 25cm (UV special coating) |
| A/D |
40M Sample/sec |
| Measurement height |
300m to 5km |
| Height resolution |
20m (lower layer) - 50m (upper layer) |
| Measured objective |
Temperature, Humidity and Aerosol |
| Measurement principle |
Temperature : Rayleigh scattering |
| Humidity : Raman scattering |
| Aerosol : Mie scattering |
| Measurement time |
Less than 5 minutes (6000 shot) |
| Software |
Measurement software, Analyze software |
|