Portable spectroradiometer MS-720 is developed to measure the spectral
irradiance in the range from Visible to Near-Infrared region. MS-720
is mainly used for the data acquisition for Geological features,
Vegetation, Oceanography, Remote sensing and others.
MS-720 is hand-heid instrument with an integrated LCD and an intemal
memory. Therefore, it's suitable for the outdoor stand-alone measurement.
A small-sized grating s~ectrometer is int~grated. Spectral irradiance
is measured by using diode array.
The data can be downloaded to PC using the attached software "720Man.exe",
and the data can be saved and edited in the PG. "720Man.exe" calculates
the integral radiation irradiance and compares the data with the
reference data.
|
|
l. Portable
MS-720 is portable (WIOO x Dl6S x H60mm. 720g) and suitable for
the outdoor measurement. The operation is easy with the integrated
LCD and the tact switch.
2. Long stability
The less aging diffuser, grating spectrometer and diode-array are
adopted in MS-720. MS-720 is durable because there is no movable
part to change the spectrum. This is accomplished by the adoption
of diode-array. MS-720 software a utomatically deterrnines the optimal
shutter opening time for the spectral measurement based on the respective
spectral intensity.
3. Substantlal software functions
Both "Catalogue View" of the plural spectral data and the "Detailed
VievV' of the single spectral data are available by MS-72C analysis
software 720Man.exe. 720Man.exe. saves the spectral data of inm
and 5nm inten/al. It also compares the data with the reference.
|
Spectrum and calculated value are shown on on-board LCD
|

Sample measurement : Irradiance permeated through a foliage plant
|
|
|
|
|
|
|
MS-720 measurement can be performed combined with
tripod
|
Aperture
(FOV180° )
|
Bottom side of MS-720
(From the left : Power switch, DC power, RS232C, USB, GPS)
|
Battery box
|
| Specificastions |
MS-720 (Hardware) |
| Wavelength range |
350 - 1,050nm |
| Wavelength interval |
3.3nm |
| Wavelength resolution |
l0nm |
| Wavelength accuracy |
Less than 0.3nm |
Aperture angle
(FUll view angle) |
180° (Option: 45°, 25°, 10°) |
| Stray light |
Less than 0.15% |
| Temperature dependency |
:±5% |
| Unit of the output |
W/m²/µm |
| Measurement time |
0.005 to 5sec/Automatic controlled time |
| Maximum number of the data |
800 (Approximately 2 minutes for download) |
| Power supply |
AA battery x 4
Optional AC adapter DC5V |
| Size |
W100 x D165 x H60 (mm) |
| Weight |
Approx. 720g (Including batteries and Aperture)
|
| Communication |
RS-232C/USB |
| Working conditions |
Temperature : -10 to +40
Humidity : 0 to 90%RH (no condensation) |
| Specificastions |
MS-720 (Software) |
| Program |
Analysis software : 720Man.exe |
| OS |
Microsoft Windows 98Second Edition/2000/XP |
| Interface |
RS-232C/USB |
| Functions |
Display the data (List view/Detailed view)
Save the data (CSV format)
Analyze the data (Data comparison) |
| Specificastions |
PC |
| CPU |
Better than Pentium 200MHZ (recommended) |
| Memory |
>=64MB (recommended) |
| Hard disc |
>=10G B (recommended)
|
| Screen |
>=l024 x 768 (recommended) |
| Others |
CD-ROM drive, RS232C port, USB port |
|